Early user experience and lessons learned using ultra-portable digital X-ray with computer-aided detection (DXR-CAD) products: A qualitative study from the perspective of healthcare providers

🔗2023

🔗Journal/Publication: PLOS one

🔗Read it in full version: https://doi.org/10.1371/journal.pone.0277843

Abstract

Background
Recent technological and radiological advances have renewed interest in using X-rays to screen and triage people with tuberculosis (TB). The miniaturization of digital X-ray (DXR), combined with automatic interpretation using computer-aided detection (CAD) software can extend the reach of DXR screening interventions for TB. This qualitative study assessed early implementers’ experiences and lessons learned when using ultra-portable (UP) DXR systems integrated with CAD software to screen and triage TB.

Methods
Semi-structured interviews were conducted with project staff and healthcare workers at six pilot sites. Transcripts were coded and analyzed using a framework approach. The themes that emerged were subsequently organized and presented using the Consolidated Framework for Implementation Research (CFIR).

Results
There were 26 interviewees with varying roles: supervisory, clinicians, radiographers, and radiologists. Participants recognized the portability as the main advantage, but criticize that it involves several compromises on throughput, internet dependence, manoeuvrability, and stability, as well as suitability for patients with larger body sizes. Furthermore, compared to using hardware and software from the same supplier and without digital health information systems, complexity increases with interoperability between hardware and software, and between different electronic health information systems. Currently, there is a limited capacity to implement these technologies, especially due to the need for threshold selection, and lack of guidance on radiation protection suitable for UP DXR machines. Finally, the respondents stressed the importance of having protected means of sharing patient medical data, as well as comprehensive support and warranty plans.

Conclusion
Study findings suggest that UP DXR with CAD was overall well received to decentralize radiological assessment for TB, however, the improved portability involved programmatic compromises. The main barriers to uptake included insufficient capacity and lack of guidance on radiation protection suitable for UP DXR.